Defibrillation-protected applied parts

Martin Bosch

29/07/2019

60601-1 defines defibrillation-protected applied parts (commonly referred to as defibrillation-resistant). Applied parts must be as defined in other articles (e.g., here) as already described. For some application parts, defibrillation resistance is required by standards (e.g., 60601-2-27 for ECG).

 

A defibrillation-protected applied part refers to the property of an applied part to function when used simultaneously with a defibrillator without impairing the function of the defibrillator.
The fulfillment of this property is tested in 60601-1 through three tests:

 

  • Common mode test

Apply a defibrillation pulse to all interconnected patient terminals against protective conductors or earth. Ensure that no hazardous voltages are present at any other outputs and connections (and accessible metallic parts).

  • Differential clock test

Apply a defibrillation pulse to each patient terminal individually, with the other patient terminals connected to ground. Again, no hazardous voltages may be present at any other outputs or terminals.

  • Energy reduction test

The energy delivered to the patient may be reduced by a maximum of 10 % by a defibrillation-protected applied part, so that the effectiveness of the defibrillation is maintained.

After all tests, the basic safety and essential performance characteristics of the tested device must still be met. The tests and corresponding test setups are described in detail in 60601-1 and will not be discussed further here. The setup is shown in principle in the following figure.

In addition, there is a requirement that the clearance and creepage distances of defibrillation-protected application parts must not be less than 4 mm.

What does the requirement for a defibrillation-protected application part mean in development?

For defibrillation-protected devices with a galvanic connection to the patient, such as ECG machines or other stimulation devices, the patient connections must be appropriately protected. There are various topologies for this protection circuit, but essentially the protection circuit consists of a series resistor for current limiting and a component for voltage limiting, such as a diode, Zener diode, or gas discharge tube.

Component specifications vary depending on the application. For example, an ECG input stage should generally have the following characteristics:

  • Series resistance in the range of 5 kOhm to 50 kOhm for current limiting

The series resistance must be specified for:

  • Pulse load
  • Dielectric strength
  • Required air and creepage distances

Bypass diodes such as MAX30034 or BAV199 specified for:

  • Pulse load
  • Electricity

Once a circuit topology has been chosen for the input circuit, it can be easily simulated with Spice. This simplifies component dimensioning, as measurements at high voltages are not so straightforward and subject to significant noise.

Ultimately, the circuit must pass the tests defined in 60601-1.

If you have any further questions on this topic, please leave us a comment or send us an email.

 

Best regards

Martin Bosch


Written by Martin Bosch

Martin Bosch is a dedicated hardware developer who pursues his passion for electronics at MEDtech Ingenieur GmbH. His expertise includes the development of embedded electronics, specifically for medical applications. His focus is on the design of printed circuit boards and circuits that integrate both microcontrollers and analog circuitry. These are used in a wide variety of devices, from blood analyzers to defibrillators.


More articles

  • 26/02/2026
  • General, hardware, quality

According to the German Resuscitation Registry 2024, 370 people per day in Germany suffer a sudden cardiac arrest. That's over 136,000 cases per year, mostly occurring outside of hospitals. ...

Read more
  • 12/02/2026
  • General, hardware, mechanics, technology, testing, validation

How are sensors for cardiopulmonary resuscitation validated? In our case, we specifically asked this question for a sensor used to measure compression depth during resuscitation. Why is a CPR feedback sensor validated? ...

Read more
  • 05/02/2026
  • General, Hardware, Requirements Engineering, Software, Zephyr

What difference can a CPR feedback system make to cardiopulmonary resuscitation? In the event of cardiac arrest, the transport of oxygen to the body's cells stops immediately, leading to irreversible damage. The cells die. ...

Read more